Four dual polarization SIS receivers are available at the telescope
for the upcoming observing season. They are designated according to
the dewar in which they are housed (A, B, C, or D), followed by the
center frequency (in GHz) of their tuning range. Their main
characteristics are summarized in Tab. 1. All receivers are
linearly polarized with the E-vectors, before rotation in the
Martin-Puplett interferometers, either horizontal or vertical in
the Nasmyth cabin. Up to four of these eight receivers can be
combined for simultaneous observations in the four ways depicted in
Tab. 1. Note that they cannot be combined with HERA nor with
the bolometers. Also listed are typical system temperatures which
apply to average
summer/winter weather (7mm/4mm of water) at the center of the tuning
range and at 45
elevation. All receivers are tuned by the
operators from the control room. Experience shows that it normally
takes not more than 15 min to tune four such receivers.
| receiver | polari- | combinations | tuning range |
|
|
remark | ||||||
| zation | 1 | 2 | 3 | 4 | GHz | K | dB | GHz | GHz | K | ||
| A100 | V | 1 | 3 | 80 - 115.5 | 60 - 80 | 1.5 | 0.5 | 120 | ||||
| B100 | H | 1 | 4 | 81 - 115.5 | 60 - 80 | 1.5 | 0.5 | 120 | 1 | |||
| C150 | V | 2 | 4 | 129 - 183 | 70 - 125 | 15 - 25 | 4.0 | 1.0 | 200 | |||
| D150 | H | 2 | 3 | 129 - 183 | 80 - 125 | 8 - 17 | 4.0 | 1.0 | 200 | |||
| A230 | V | 1 | 3 | 197 - 266 | 85 - 150 | 12 - 17 | 4.0 | 1.0 | 450 | 2 | ||
| B230 | H | 1 | 4 | 197 - 266 | 95 - 160 | 12 - 17 | 4.0 | 1.0 | 450 | 2 | ||
| C270 | V | 2 | 4 | 241 - 281 | 125 - 250 | 10 - 20 | 4.0 | 1.0 | 1000 | 3 | ||
| D270 | H | 2 | 3 | 241 - 281 | 150 - 250 | 9 - 13 | 4.0 | 1.0 | 1000 | 3 | ||
| HERA1 | H | 217 - 268 | 100 - 380 | 4.0 | 1.0 | 400 | 2, 4 | |||||
| HERA2 | V | 216 - 242 | 100 - 350 | 4.0 | 1.0 | 400 | 2, 4 | |||||